Electron Microscopy Studio · Reference
FFT
Applies to Electron Microscopy Studio 2.1 and later
FFT shows the Fourier transform of an image or of a region of it, measures its spots, and filters the image through a mask. This page lists every control of the part, which is in the Imaging workspace.
Region
| Control | What it does |
|---|---|
| Transformed | What is transformed: the whole image, or the region with its position and size |
| Use Centred Square | Sets the largest centred square as the region |
| Clear Region | Transforms the whole image again |
| Δq, or Δq (x) and Δq (y) | The frequency step between neighbouring bins of the transform |
Draw a region on the image to transform only that part.
Before the transform
| Control | Choices | What it does |
|---|---|---|
| Window | Tapers the edges of the region before the transform. A hard edge streaks the spectrum along both axes. | |
| Smoothing | None, Gaussian with σ | Smooths the region first |
| Band-pass | None, Difference of Gaussians, Butterworth high-pass, Butterworth low-pass | Keeps a band of frequencies. Low σ or Cutoff sets it. |
The inspector sums the steps up in one line that starts "Before the transform:". The picture is drawn as log₁₀ of the power. The window, the smoothing and the band-pass change the transform the spots are measured on: they change a spot's power, not its position.
Spots
Click a spot on the transform to measure it. The spot is listed in the inspector with its values.
| Control | What it does |
|---|---|
| Find Spots | Puts the strongest reflections on the transform, each measured as a click would be |
| Threshold, Distance, Most spots | The finder's settings: the level as a fraction of the range, the least distance between spots in bins, and the most spots it places |
| Measure at… | Measures at a position you type, kx and ky in bins |
| In Mask | Passes this spot in the masked inverse. Only a measured spot can be in the mask. |
| Remove, Remove All | Deletes one spot, or all |
A yellow ring is a measured spot, a blue one the selected spot, and a dashed one a spot that could not be measured. ⚠ marks a spot with another peak close by: its position is less certain.
Radial Profile
The mean power against |q|. Point at the chart to read a bin.
Masked Inverse
Appears once a spot is measured. The inverse transform through a mask around the spots marked In Mask shows where in the image those frequencies come from.
| Control | Unit | What it does |
|---|---|---|
| Radius | bins | The mask passes everything within this distance of the centre of a spot |
| Edge | bins | The mask falls to nothing over this distance beyond the radius |
HRTEM Filters
A re-implementation of D. R. G. Mitchell's HRTEM filter script for Digital Micrograph (ABSF and Wiener). Parameters holds the settings, Filter runs it, and Last run says what the run did.
Filtered images are for looking, not for measuring
The app states the limits itself. A lattice that falls between two Fourier bins loses amplitude: on synthetic lattices it keeps 5 to 57 % in a 128 px region and 77 to 97 % at 256 px. Intensities are not preserved: the mean of the output is below that of the input.
Display
View chooses what the image pane shows, for example the image itself or the masked inverse.