4DSTEM Studio

4DSTEM Studio · Reference

Imaging

Applies to 4DSTEM Studio 5.4 and later

Imaging turns the scan into pictures: a virtual image from a detector you draw, and the diffraction of a region you choose. This page lists every control of its two tasks.

Virtual imaging

Control What it does
Direction What dragging on the panes controls. Detector: the virtual detector on the diffraction pane, which produces the image of the scan. Region: the scan region, which produces the diffraction pattern.

Detector → real space

Control What it does
Shape The detector drawn on the diffraction pane: Circle, Annulus, Rectangle or Point (a single pixel)
Preset BF (bright field), ADF (annular dark field) and HAADF (high-angle annular dark field) set the detector in one click. Custom is shown once you change it.
Centre and radii The detector's position and size in detector pixels. Type them, or drag the detector on the diffraction pane.

The image of the scan updates while you drag. Each pixel of it is the signal that fell on the detector at that scan position.

Without a measured beam the presets are fractions of the detector. Once Calibrate Origin in Prepare has measured the probe radius r, BF becomes the disk of radius r and ADF the annulus from 3r to 6r, as in py4DSTEM's virtual-imaging tutorial. HAADF stays a fraction of the detector. The presets are not set in mrad: to report collection angles, read the radii against the Q calibration.

Region to diffraction

Control What it does
Shape The scan region summed into the diffraction pattern: one position, a rectangle or a circle
Half-width or Radius The size of the region in scan pixels

With a single position, dragging on the image steps through the patterns. With a rectangle or a circle, the diffraction pane shows the summed pattern of the region and updates while you drag.

Diffraction groups

This task sorts the scan positions into groups with similar diffraction patterns. It needs no calibration and no model of the sample, so it is a quick first look at what a scan contains.

Control Choices What it does
Binned size 8 × 8, 16 × 16, 32 × 32 Each pattern is log-scaled, normalised and binned to this many pixels per side before grouping
Components a number The principal components kept from the binned patterns
Groups a number The number of groups (k-means on the principal components)
Group Patterns Runs the grouping over every scan position
Find similar to current position Shows how similar every pattern is to the one at the selected scan position, using the last grouping

After a run the section reports:

  • how much of the variance the first three components explain,
  • the size of each group,
  • the settings the result was computed with, including the seed, so the run can be repeated exactly.

"From an earlier run — the settings above have changed" means the map on screen no longer matches the settings. Run Group Patterns again.

What the groups are

The groups are regions with similar patterns, not phases. Two orientations of one phase fall into different groups, and thickness changes can split one. Use the groups to choose where to look, then identify what is there in Bragg Disks and Crystal Maps.

Method

Virtual imaging follows py4DSTEM: Savitzky et al. (2021), Microsc. Microanal. 27, 712. The grouping is principal component analysis followed by k-means. Its input, log-scaled, normalised and binned patterns, has no counterpart in py4DSTEM, which groups maps you supply; the principal components are checked against scikit-learn. The numbers are in the app's parity notes.